Volume Range μL | Increment μL | TestVolume uL | accuracy error | precision error | ||
uL | % | s.d.uL | CV% | |||
0.5-10uL | 0.01uL | 10 1 | ±0.10 ±0.035 | ±1.0 ±3.5 | 0.05 0.03 | 0.5 3.0 |
5-50uL | 0.1uL | 50 5 | ±0.40 ±0.15 | ±0.8 ±3.0 | 0.15 0.125 | 0.3 2.5 |
30-300uL | 1uL | 300 30 | ±1.8 ±0.9 | ±0.6 ±3.0 | 0.6 0.21 | 0.2 0.7 |
100-1000uL | 5uL | 1000 100 | ±6.0 ±3.0 | ±0.6 ±3.0 | 2.0 0.6 | 0.2 0.6 |
Volume Range μL | Increment μL | TestVolume uL | accuracy error | precision error | ||
uL | % | s.d.uL | CV% | |||
0.5-10μl | 0.01μl | 10 | ±0.20 | ±2.0 | 0.1 | 1.0 |
1 | ±0.08 | ±8 | 0.05 | 5.0 | ||
10-100μl | 0.1μl | 100 | ±0.8 | ±0.8 | 0.3 | 0.3 |
10 | ±0.3 | ±3 | 0.2 | 2 | ||
30-300μl | 1μl | 300 | ±1.8 | ±0.6 | 0.9 | 0.3 |
30 | ±0.9 | ±3 | 0.3 | 1.0 |